| Model Number |
6935M49 |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Shock from Patient Lead(s) (3162)
|
| Date of Event |
03/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead has t-wave oversensing (twos) post ventricular pacing.The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the rv lead delivered a shock post twos.The lead was reprogrammed.
|
| |
|
Search Alerts/Recalls
|