• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE MEDIAL DOME PAT 29MM; ATTUNE IMPLANT : KNEE PATELLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE MEDIAL DOME PAT 29MM; ATTUNE IMPLANT : KNEE PATELLA Back to Search Results
Model Number 1518-20-029
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 03/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the sales rep was notified that the patient had an infected left knee.The surgical removal of implants was planned for on an urgent basis on this day.Surgery was performed implants were removed and a prostalac knee spacer was implanted.Doi: (b)(6) 2018, (b)(6) 2019 (insert).Dor: (b)(6) 2020, left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Per internal procedures, the event information and any investigational inputs received as part of required follow up were reviewed.For this investigation, no immediate action was required as no alleged deficiency with the device was identified.Monitor ¿ exempt per wi-7915 - known possible complication of joint replacement surgery.No device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample associated with this report, it was not possible to determine if the device failed to meet specifications at the time it was released for distribution.The device associated with this event was used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No evidence was found indicating product error was a contributing factor.The need for corrective action was not indicated.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot null device history batch null device history review null.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE MEDIAL DOME PAT 29MM
Common Device Name
ATTUNE IMPLANT : KNEE PATELLA
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9907895
Report Number1818910-2020-09803
Device Sequence Number690146
Product Code JWH
UDI-Device Identifier10603295056669
UDI-Public10603295056669
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1518-20-029
Device Catalogue Number151820029
Device Lot Number8583168
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/16/2020
Supplement Date Received by Manufacturer04/27/2020
Initial Report FDA Received Date03/31/2020
Supplement Report FDA Received Date04/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 6 8MM; ATTUNE CR FEM LT SZ 6 CEM; ATTUNE FB TIB BASE SZ 5 CEM; ATTUNE MEDIAL DOME PAT 29MM; SMARTSET GHV GENTAMICIN 40G; SMARTSET GHV GENTAMICIN 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-