| Model Number |
9-VSDMPIHDE-024 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Myocardial Infarction (1969); Heart Failure (2206)
|
| Date of Event |
03/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 24mm amplatzer post infarct vsd occluder in a three day old myocardial infarction.The device was mis-sized too small and pulled through.The patient was put on extracorporeal membrane oxygenation.The defect is planned to be surgically fixed when a bypass surgery is performed on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information to b5.Correction to: b1, b2, h1, h2, and h10.Manufacturer report 2135147-2020-00141, should not have been reported as a medical device report (mdr) as the event is not reportable.Upon review, the amplatzer post infarct vsd occluder should not have been submitted as a medical device report (mdr).There is no indication that abbott product caused a serious injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 24mm amplatzer post infarct vsd occluder in a three day old mi.The device was mis-sized too small and pulled through.The patient was put on extracorporeal membrane oxygenation (ecmo).There was no allegations to the device that caused the patient to be put on ecmo and no complications due to the procedure or device.The defect is planned to be surgically fixed when a bypass surgery is performed on the patient.
|
| |
|