| Model Number |
DSA-135-01 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the distal portion of the cannula broke off and remained in patient.The procedure was attempted with a 135 degree catheter using a duct that had an estimated size of about 12-14 mm and ueus scope.The needle tip went in and there were some reduced visibility after the stylet was removed which they thought it was relatively normal or expected.The contrast was injected and there were some leakage of contrast outside the duct which was likely from the top fenestration.The stylet was pushed back in for position adjustment but there was resistance from about 10 cm before it would have been hubbed.The attempted adjustment seemed unsuccessful so the physician decided to pull the cannula back into the scope.After that, the stylet was successfully reinserted.The device was removed from the scope and they noticed an abnormal appearance of the distal tip.Approximately 90-95% of the fenestrated end sheered off into the duct with a somehow horseshoe shape.The customer stated that motions were careful with the scope and there was not much rotation of the cannula during the procedure, just a mm or two.The physician was uncertain as to what point of the procedure the detachment occurred.The patient was not reimaged after the procedure and was sent to a different facility for further work-up/treatment.The physician seemed to think that there was likely some bile leakage and the patient needed decompression of the biliary system.There was post procedural pain noted and the patient was given antibiotics.They also performed ercp (endoscopic retrograde cholangiopancreatography) to the patient.Based on the pre-ercp imaging, the broken portion of the catheter appeared to be partially within the duct but the majority was outside of the duct.There was free fluid outside of the duct compatible with biliary leak.They performed a percutaneous drainage and based on the ercp fluoroscopy/balloon sweep of the duct, the broken portion of the cannula was not identified within the duct.It was expected that the patient will be inpatient for somewhere around 3-5 days, and that direct visualization with a biliary camera may be performed at a follow-up ercp.The pre-procedure diagnosis was chronic abdominal pain, abnormal lst, dilated cbd (common bile duct).The name of the procedure performed to the patient was eus guided rendezvous.The removal of the needle was anticipated, the tip of the needle was still in the patient's biliary duct and the patient was aware of it.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g4, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was returned for evaluation.Visual inspection identified that approximately an inch of the needle distal end had broken off.It was reported that the needle tip had broken off into the patient.A related device issue was confirmed.A preliminary root cause is likely manufacturing.The investigation is ongoing to determine final root cause.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the distal portion of the cannula broke off and remained in patient.The procedure was attempted with a 135 degree catheter using a duct that had an estimated size of about 12-14 mm and ueus scope.The needle tip went in and there were some reduced visibility after the stylet was removed which they thought it was relatively normal or expected.The contrast was injected and there were some leakage of contrast outside the duct which was likely from the top fenestration.The stylet was pushed back in for position adjustment but there was resistance from about 10 cm before it would have been hubbed.The attempted adjustment seemed unsuccessful so the physician decided to pull the cannula back into the scope.After that, the stylet was successfully reinserted.The device was removed from the scope and they noticed an abnormal appearance of the distal tip.Approximately 90-95% of the fenestrated end sheered off into the duct with a somehow horseshoe shape.The customer stated that motions were careful with the scope and there was not much rotation of the cannula during the procedure, just a mm or two.The physician was uncertain as to what point of the procedure the detachment occurred.The patient was not reimaged after the procedure and was sent to a different facility for further work-up/treatment.The physician seemed to think that there was likely some bile leakage and the patient needed decompression of the biliary system.There was post procedural pain noted and the patient was given antibiotics.They also performed percutaneous retroperitoneal (extraluminal) drain to the patient.The extraluminal drain was removed maybe about a week after it had been placed.Based on the pre-ercp imaging, the broken portion of the catheter appeared to be partially within the duct but the majority was outside of the duct.There was free fluid outside of the duct compatible with biliary leak.They performed a percutaneous drainage and based on the ercp fluoroscopy/balloon sweep of the duct, the broken portion of the cannula was not identified within the duct.It was expected that the patient will be inpatient for somewhere around 3-5 days, and that direct visualization with a biliary camera may be performed at a follow-up ercp.The pre-procedure diagnosis was chronic abdominal pain, abnormal lst, dilated cbd (common bile duct).The name of the procedure performed to the patient was eus guided rendezvous.The removal of the needle was anticipated, the tip of the needle was still in the patient's biliary duct and the patient was aware of it.A repeat ercp was not yet performed but it was anticipated in a few months.Eus and spyglass procedures are also anticipated.
|
| |
|