MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Intermittent Continuity (1121); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow up-report.
 
Event Description
It was reported that the device stopped ventilating during use.There was no patient injury reported.
 
Manufacturer Narrative
The responsible dräger service engineer could confirm the reported issue during on-site checking and trace it back to the ventilator motor.The entire motor assembly was replaced, consequently.Unfortunately, neither the replaced parts nor the logfile were available for in-depth investigation.According to the reported error code 14.1499, the device forced a shutdown of automatic ventilation due to a detected wrong motor position.The motor speed is monitored continuously; speed fluctuations caused e.G.By an abraded collector disc will result in a deviation between measured and expected piston position.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes based on the investigation, that the device behaved as specified upon the malfunction of a single component; no patient consequences have been reported.The repair exchange of the motor unit has fully solved the device problem.The number of similar cases is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that the device stopped ventilating during use.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key9909457
MDR Text Key185914023
Report Number9611500-2020-00106
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-