Additional information: d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the loading unit had a full staple complement.The anvil clamping mechanism was deformed.The tissue stop on the left side of the loading unit was peeled back towards the distal end.Due to the condition of the loading unit functional testing could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the tissue stop deformation may occur when the tissue stop catches on the trocar during removal, or if an attempt is made to remove the device while in the articulated position.Any forceful effort to remove the loading unit will result in the tissue stop being peeled distally and prevent the loading unit from being removed from the trocar.Replication of the bowed anvil, buckled knife bar assembly, and deformed anvil clamping feature may occur under the following conditions: 1.Applica tion over tissue that is beyond the recommended thickness range.2.Application with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.Damage to the anvil by any of the previously mentioned methods may cause the tissue stops to begin to flare outwards as the anvil deforms.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|