• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAS
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
 
Event or Problem Description
Information indicated by medtronic that the insulin pump had a motor error.Customer stated that the drive support cap was sticking out.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9909773
Report Number3004209178-2020-92786
Device Sequence Number2319848
Product Code OYC
UDI-Device Identifier00613994402097
UDI-Public(01)00613994402097
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAS
Device Catalogue NumberMMT-522NAS
Device Lot NumberA5522NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Initial Date Received by Manufacturer 03/23/2020
Supplement Date Received by Manufacturer06/02/2020
Initial Report FDA Received Date04/01/2020
Supplement Report FDA Received Date06/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight152
-
-