• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAP
Medical Device Problem Codes Mechanical Problem (1384); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had motor error alarm often and also had a crack at the top facing.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose and protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test due to motor error alarm.Device received with cracked case from display window corner and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9909961
Report Number3004209178-2020-92823
Device Sequence Number19676752
Product Code OZO
UDI-Device Identifier00643169404335
UDI-Public(01)00643169404335
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAP
Device Catalogue NumberMMT-751NAP
Device Lot NumberA2751NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Initial Date Received by Manufacturer 03/23/2020
Supplement Date Received by Manufacturer06/05/2020
Initial Report FDA Received Date04/01/2020
Supplement Report FDA Received Date06/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
Patient Weight225
-
-