MAUDE Adverse Event Report: BIOMET MICROFIXATION STERNALOCK BLU SYSTEM SCREW, CANCELLOUS CROSS-DRIVE LOCKING 2.4 X 18MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION STERNALOCK BLU SYSTEM SCREW, CANCELLOUS CROSS-DRIVE LOCKING 2.4 X 18MM; PLATE, FIXATION, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 03/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00190, 0001032347-2020-00191.Concomitant medical products: sternalock blu system plate, 8 hole x, part# 73-2623, lot# ni.Sternalock blu system screw, cancellous cross-drive locking 2.4 x 16mm, part# 73-2416, lot# ni.Sternalock blu system screw, cancellous cross-drive locking 2.4 x 18mm, part# 73-2418, lot# ni.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source ¿ (b)(6).
 
Event or Problem Description
It was reported the patient underwent a revision two (2) weeks following implantation of sternal closure plates and screws due to mechanical sternal dehiscence.The implant failure was discovered upon follow-up after mitral valve replacement.The sternal closure implants were removed in the revision and a redo sternal closure was performed.No additional patient consequences were reported.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed because a revision surgery was reported.The devices were not returned for investigation and no photos, scans, x-rays, or physician's reports were provided.For these reasons, no functional testing or inspections could be performed.The dhr could not be reviewed due to the lot number remaining unknown.There are no indications of manufacturing defects.For this part (73-2418) in the previous one year (from the notification date), the full complaints history review yielded a complaint rate of 0.14%.The most likely underlying cause of the complaint is due to the patient condition.There are no indications of manufacturing defects.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERNALOCK BLU SYSTEM SCREW, CANCELLOUS CROSS-DRIVE LOCKING 2.4 X 18MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9910378
Report Number0001032347-2020-00192
Device Sequence Number14291685
Product Code HRS
UDI-Device Identifier00841036147587
UDI-Public00841036147587
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K110574
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number73-2418
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/20/2020
Supplement Date Received by Manufacturer09/28/2020
Initial Report FDA Received Date04/01/2020
Supplement Report FDA Received Date09/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-