| Model Number |
E100-31M-00 |
| Medical Device Problem Code |
Perivalvular Leak (1457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/23/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 31mm epic valve was selected for implant.The valve was fully implanted and the surgeon elected to pressure test the valve while the patient was on bypass.The surgeon was not comfortable with the way the valve looked; it was reported there was leak when the ventricle was pressurized.The valve was intra-operatively explanted and a new 31mm epic valve was successfully implanted.It was reported that there was a clinically significant prolonged procedure time due to the exchange.No patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the surgeon not being comfortable with the way the valve looked and a leak when the ventricle was pressurized was reported.No anomalies were found with the valve cusps, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper cuspal coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
|
| |
|