| Model Number |
NSEAL545RH |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # t94m0u.Investigation summary: the device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The jaws were found attached to the instrument.In addition no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the device did not meet the customer¿s expectations.It is unknown how the case was completed.It is unknown if there were any adverse consequences to the patient.
|
| |
|