• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE CD; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE CD; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 03/06/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Udi: (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during sterilization trial broke.No patient impact.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination exhibits signs of repeated use and has fractured on the lateral side of the post all pieces returned reference attached photographs.Device history record (dhr) was reviewed and no discrepancies identified.A definitive root cause cannot be determined.Evaluation of the returned device identified the fracture was consistent with the tasp fractures analyzed in a zrm, which identified that the common failure modes for the tasp devices include either bending overload or low cycle fatigue culminating in bending overload as evident by the presence of hackle marks, river lines and striations features on the fracture surface.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE CD
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9912411
MDR Text Key190040822
Report Number0001822565-2020-01252
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024244962
UDI-Public(01)00889024244962
Combination Product (y/n)N
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 05/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527000303
Device Lot Number64273921
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-