| Model Number |
7213 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 85% stenosed target lesion was located in the averagely tortuous and severely calcified mid right coronary artery.After a staged rotablator, the lesion was dilated with an emerge balloon catheter without issues, but upon removal from the patient's body the shaft then snapped 1/3 from the hub inside the 8fr guide catheter.The wire was pulled back into the guide catheter and snared the broken shaft without removing the guide catheter.The lesion then dilated with a different balloon catheter.A 2.50 x 32 synergy drug-eluting stent was advanced but failed to cross and so another balloon was used to pre-dilate the lesion.The 2.50 x 32 stent was advanced for the second time but the shaft broke at the mid portion of the catheter while advancing into the guide.The stent was pulled out from the guide catheter without removing the device from it.The procedure was completed with a different device.No patient complications were reported and the patient remained stable all throughout the procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 85% stenosed target lesion was located in the averagely tortuous and severely calcified mid right coronary artery.After a staged rotablator, the lesion was dilated with an emerge balloon catheter without issues, but upon removal from the patient's body the shaft then snapped 1/3 from the hub inside the 8fr guide catheter.The wire was pulled back into the guide catheter and snared the broken shaft without removing the guide catheter.The lesion then dilated with a different balloon catheter.A 2.50 x 32 synergy drug-eluting stent was advanced but failed to cross and so another balloon was used to pre-dilate the lesion.The 2.50 x 32 stent was advanced for the second time but the shaft broke at the mid portion of the catheter while advancing into the guide.The stent was pulled out from the guide catheter without removing the device from it.The procedure was completed with a different device.No patient complications were reported and the patient remained stable all throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 48cm distal of the strain relief.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported broken shaft as the hypotube was separated.
|
| |
|
Search Alerts/Recalls
|