| Model Number |
1012-14-020 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Test Result (2695); No Code Available (3191)
|
| Date of Event |
10/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Unf and medical records received.After a review of the medical record, the patient was revised to address failed right total hip replacement secondary to metallosis.Revision operative notes reported that metallosis was noted in the hip capsule.Patient clinic visit note on (b)(6) 2019, reported a high cobalt level.Doi: (b)(6) 2009; dor: (b)(6) 2019; right hip.Please see (b)(4) for the left hip revision.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|