| Model Number |
82885 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.A visual and functional inspection revealed that a catheter and coil were returned for this complaint.The coil was stuck inside the catheter.It was necessary to cut the catheter in order to release the coil.The main coil was found stretched and kinked.No more damages were found in the device.During the product analysis the investigator was unable to remove the coil from the catheter due to the stretched and kinked.The investigator cut off the catheter and successfully removed the coil.A microscopic and dimensional inspection revealed that the proximal zap tip and distal tip of the main coil was within its specification aside from the fiber bundles which was missing due to the coil stretching.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 26mar2020.It was reported that the coil got stuck in the catheter.The target lesion was located in the portal vein.A 7mmx40mm 2d helical - 35 embolic coil was selected for use.During procedure, it was noted that the coil got stuck inside the delivery catheter.The device was removed as a unit and another of the same device was used to complete the procedure.No patient complications were reported and patient status was good.However, device analysis revealed missing fiber bundles.
|
| |
|