| Model Number |
CQF7588 |
| Medical Device Problem Codes |
Deflation Problem (1149); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the device is pending.The investigation of the reported event is currently underway.(expiry date: 08/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the brachial artery, the pta balloon allegedly could not be fully deflated.It was further reported that a needle stick was used to percutaneously deflate the balloon.The balloon was removed from the patient and another device was used to complete the procedure.There were no further reported complications to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a lot history review was conducted and it was determined that a device history record (dhr) review was not required.Investigation summary: the device was returned for evaluation.A visual inspection found bunching to the outer catheter shaft.Therefore, the investigation is confirmed for the identified outer catheter shaft bunching.Due to the condition of the returned device, functional testing was unable to be performed.Therefore, the investigation is inconclusive for the reported deflation issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 08/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the brachial artery, the pta balloon allegedly could not be fully deflated.It was further reported that a needle stick was used to percutaneously deflate the balloon.The balloon was removed from the patient and another device was used to complete the procedure.There were no further reported complications to the patient.
|
| |
|