According to the reporter, during a septorhinoplasic procedure, on wound closure, just when the surgeon throw the suture through the tissue, the needle detached from the thread on the four devices.They used another suture to complete the case.There was no patient injury.
|
|
Additional information: d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of five devices.The visual inspection of the sample noted five sutures and five needles.Four of these needles were detached from their respective sutures.Upon microscopic inspection, instrument marks were noted near the swaged end of each disengaged needle.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.The needle should always be held in the middle where the diameter is greatest.Grasping the needle near the swaged end could cause bends or breaks, or damage the connection between the needle and suture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|