U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10607
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that stent damage occurred.A 3.00 x 38 synergy drug- eluting stent was selected for use.However, during unpacking, it was noted that the stent was deformed.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device is a combination product.Device evaluated by mfr.: synergy ii us mr 3.00 x 38 mm stent delivery system was returned for analysis.An examination of the crimped stent found stent damage.Stent struts from the distal region of the stent were noted to be lifted from their crimped position and pulled.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of distal tip damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that stent damage occurred.A 3.00 x 38 synergy drug- eluting stent was selected for use.However, during unpacking, it was noted that the stent was deformed.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9914676
Report Number2134265-2020-04288
Device Sequence Number2742076
Product Code NIQ
UDI-Device Identifier08714729840411
UDI-Public08714729840411
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2021
Device Model Number10607
Device Catalogue Number10607
Device Lot Number0024739159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2020
Initial Date Received by Manufacturer 03/19/2020
Supplement Date Received by Manufacturer05/01/2020
Initial Report FDA Received Date04/02/2020
Supplement Report FDA Received Date05/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-