MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1219-32-152
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Adhesion(s) (1695); Joint Dislocation (2374); No Code Available (3191)
Date of Event 12/13/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Unf and medical records received.After review of medical records, the patient was revised for recurrent dislocation.It was also reported that x-ray found that the patient had a significant positioning of the left hip.Operative notes reported that there was a fractured part on the tip of the greater trochanter.The fracture fragment was caught with mersilene tape several times and it was then pulled down around the greater trochanter.It was observed that the hip joint was positioning easily and it was dislocating anteriorly.Scar tissues was making a lever to the femoral neck, and it was helping anterior dislocation.Doi: (b)(6) 2007; dor: (b)(6) 2007; left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 32IDX52OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9915807
Report Number1818910-2020-09986
Device Sequence Number19436910
Product Code LPH
UDI-Device Identifier10603295014058
UDI-Public10603295014058
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2007
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-32-152
Device Catalogue Number121932152
Device Lot NumberBH2BE1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/18/2020
Supplement Date Received by Manufacturer05/07/2020
Initial Report FDA Received Date04/02/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-