| Model Number |
B2040-100 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that a 4x100mm armada 35 percutaneous transluminal angioplasty (pta) was received by the customer with blood on the outside packaging.The device was not used and there was no patient involvement.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the reported complaint was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Based on the information reviewed and analysis of the returned device, the reported contamination appears to be due to case circumstances.The noted contamination (blood) was likely due to handling of the packaging at the account.The noted crushed portion on the bottom right-hand corner of the chipboard box may have occurred during transport for return of the device.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|