MEDTRONIC HEART VALVE DIVISION HANCOCK; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number HC15022
Medical Device Problem Codes Gradient Increase (1270); Obstruction of Flow (2423)
Health Effect - Clinical Code Pulmonary Insufficiency (2022)
Date of Event 03/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 9 years and 6 months post implant of this 22mm pulmonary valved conduit in a (b)(6)-year-old pediatric patient, there was mild pulmonary insufficiency, however the leaflets were functioning appropriately.The 22mm conduit was measured to be 19mm with a subvalvar obstruction which was thought to be a muscle bundle.The physician thought that since there was a complex lesion, distal to the conduit, primary intervention was necessary, and several balloon dilations of the lpa were performed, which relieved the gradient, however after 8 hours of complex intervention, the decision was made to delay conduit intervention since the gradient decreased to around 25mmhg, and the conduit remains implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANCOCK
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key9916784
Report Number2025587-2020-01088
Device Sequence Number2333626
Product Code LWR
UDI-Device Identifier00681490908818
UDI-Public00681490908818
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P790007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/22/2015
Device Model NumberHC15022
Device Catalogue NumberHC15022
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2020
Initial Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age19 YR
Patient Weight64
-
-