• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3548007
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Granuloma (1876)
Date of Event 02/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: not applicable.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent a breast augmentation revision with 800cc mentor memorygel breast implants and experienced postoperative right-sided rupture.The patient reported she underwent a sonogram and mri which identified that loose silicone from the rupture migrated to the lymph nodes.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
 
Additional Manufacturer Narrative
On april 5, 2020, a manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
On august 24, 2020, mentor received the following additional information via fda mw5095028, the date of event was on (b)(6) 2020, the patient was part of a study group for mentor in 1998, and the patient also reported that the implants are now leaking into their body and into their lymph nodes.The complications regarding the left breast implant will be reported under mrn: 1645337-2020-11413.Manufacturer¿s reference number: (b)(4).This medwatch form is for the right breast prosthesis.
 
Additional Manufacturer Narrative
On november 15, 2021, mentor received additional information indicating that the patient has been scheduled for implant explantation surgery on (b)(6) 2021 and that the deflation on the right breast has also been confirmed.On december 3, 2021, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: granuloma, material rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2021, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the gel siltex mod-rnd 800cc breast implant.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture status identified by mri evaluation includes both suspected ruptures, which are those ruptures identified by mri but not confirmed by explantation and examination of the device, and confirmed ruptures, which are those ruptures that are confirmed by evaluation of the explanted devices.The event described could not be confirmed as the breast implant was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Granulomas are noncancerous nodules that are a common tissue reaction to the presence of a variety of foreign materials, such as silicone.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
949789-8687
MDR Report Key9917072
Report Number1645337-2020-04060
Device Sequence Number14155070
Product Code FTR
UDI-Device Identifier00081317000778
UDI-Public00081317000778
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1998
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2002
Device Model Number3548007
Device Catalogue Number3548007
Device Lot Number165768
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/09/2020
Supplement Date Received by Manufacturer04/05/2020
08/24/2020
11/15/2021
12/14/2021
Initial Report FDA Received Date04/02/2020
Supplement Report FDA Received Date04/09/2020
09/11/2020
12/09/2021
12/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/1997
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-