(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use states: if resistance is felt, determine the cause before proceeding.Continuing to advance or retract the catheter while under resistance may result in damage to the vessels and / or damage / separation of the catheter.The investigation determined the reported shaft separation appears to be related to user error.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|