• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TAPERFILL FEMORAL HIP STEM; TAPERFILL HIP STEM, STANDARD, SIZE 8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TAPERFILL FEMORAL HIP STEM; TAPERFILL HIP STEM, STANDARD, SIZE 8 Back to Search Results
Model Number 425-96-008
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to the patient having a prosthesis fracture.The previous surgery and the surgery detailed in this event occurred 14 days apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available.The revision surgery was completed as intended and without incident.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to prosthesis fracture.There were no findings during this evaluation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - due to patient having a prosthesis fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERFILL FEMORAL HIP STEM
Common Device Name
TAPERFILL HIP STEM, STANDARD, SIZE 8
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key9917985
Report Number1644408-2020-00303
Device Sequence Number2818416
Product Code LPH
UDI-Device Identifier00888912078535
UDI-Public(01)00888912078535
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/13/2025
Device Model Number425-96-008
Device Catalogue Number425-96-008
Device Lot Number187R1172
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2020
Initial Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-