This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 2, 2020.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 11, 3331, 4114, 213, 67, 4315).Method code #1: 11 - testing of device from same lot/batch retained by manufacturer.Method code #2: 3331 - analysis of production records.Method code #3: 4114 - device not returned.Results code: 213 - no device problem found.Conclusions code #1: 67 - no problem detected.Conclusions code #2: 4315 - cause not established.The affected sample was not returned.However, a representative sample from the same lot number was tested for gas transfer performance per standard test method, built into a circuit with bovine blood conditioned to 37*c and hb of 12.7g/dl.The sample was determined to meet factory specifications.Root cause cannot be determined as the sample was not returned and the issue was not able to be recreated in the laboratory.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
|
Additional information received that, after changing with the oxygenator, with the flow diagram of fio2 and gas flow, shows 10l/min and with 100% fio in the blood gives a co2 of 47mmhg and po2 of 111mmhg.They perceived this as insufficient oxygenation with these gas settings.
|