Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
An instrument that was missing ball bearings was received via the worn instrument return program.There was no known patient involvement.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Returned instrument exhibits signs of repeated use (nicked or gouged) and has a missing ball bearing.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|