• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MOTOR LEGEND EHS STYLUS REFURB; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MOTOR LEGEND EHS STYLUS REFURB; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number EM200-R
Medical Device Problem Codes Overheating of Device (1437); Loss of Power (1475); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code Chemical Exposure (2570)
Date of Event 03/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusion can be drawn for the em200.Evaluation could not be performed because the motor seized and was noisy.No conclusion can be drawn for the ad03.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.The preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use for approximately 43 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that a perforator was used together with an ad03 and em200-r during a brain tumor resection procedure.They noticed that the number of rotations only reached about 4000 and an error code 16 was displayed.At that time, overheating also occurred.There was no backup drill that could be used and the procedure was discontinued for 90 minutes.During that time, a non-medtronic drill was borrowed from a different facility to complete the procedure.On follow up, it was reported that the perforator used was a codman and it was discarded.It was also added that the patient had no further issues post-surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR LEGEND EHS STYLUS REFURB
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key9920551
Report Number1625507-2020-00045
Device Sequence Number16968209
Product Code HBE
UDI-Device Identifier00613994422002
UDI-Public00613994422002
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K012456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM200-R
Device Catalogue NumberEM200-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2020
Initial Report FDA Received Date04/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-