| Model Number |
MMT-523LNAB |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the reservoir compartment cracked and chipped off.Customer stated reservoir was unable to lock in place when inserted into insulin pump.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with broken reservoir tube lip.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.
|
| |
|