Model Number 9733560XOM |
Device Problems
Mechanical Problem (1384); Output Problem (3005)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: other relevant device(s) are: product id: 9736119r, serial/lot #: unk.
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used outside of a procedure.
It was reported that the system became unresponsive twice.
The system was hard rebooted and the system booted into a grub menu that displayed, 'grub menu, please wait.
" the representative attempted to press escape and the issue did not resolve.
|
|
Manufacturer Narrative
|
Serial number of concomitant medical products is: (b)(4).
Additional information) a medtronic representative went to the site to test the system at the time of the reported event.
Rebooting the system resolved the issue.
The system performed as intended and no parts were replaced.
The system then passed a system checkout.
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.
Rebooting the computer resolved the issue and there have been no more issues since.
|
|