| Model Number |
M0062251010 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a uromax ultra dilation balloon catheter was used in the ureter during a right ureteroscopy (urs) procedure performed on (b)(6) 2020.According to the complainant, during preparation, a leak was noted on the balloon.The procedure was completed with another uromax ultra dilation balloon catheter.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: problem code 1354 captures the reportable event of balloon leaked.Block h10: investigation result: a visual examination of the complaint device revealed the balloon had been inflated and was not refolded.A microscopic examination identified a longitudinal tear.No issues were noted on the balloon material, tip, markerbands and catheter.This failure is likely due to an interaction between the user and device, or sample, that caused or contributed to the error.It is likely that the failure may be related to unintended inappropriate use of the device and incorrect sample preparation.Therefore, the most probable root cause is unintended use error caused or contributed to event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the directions for use (dfu)/product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a uromax ultra dilation balloon catheter was used in the ureter during a right ureteroscopy (urs) procedure performed on (b)(6) 2020.According to the complainant, during preparation, a leak was noted on the balloon.The procedure was completed with another uromax ultra dilation balloon catheter.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|