• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION UROMAX ULTRA; DILATOR, CATHETER, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION UROMAX ULTRA; DILATOR, CATHETER, URETERAL Back to Search Results
Model Number M0062251010
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/13/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a uromax ultra dilation balloon catheter was used in the ureter during a right ureteroscopy (urs) procedure performed on (b)(6) 2020.According to the complainant, during preparation, a leak was noted on the balloon.The procedure was completed with another uromax ultra dilation balloon catheter.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: problem code 1354 captures the reportable event of balloon leaked.Block h10: investigation result: a visual examination of the complaint device revealed the balloon had been inflated and was not refolded.A microscopic examination identified a longitudinal tear.No issues were noted on the balloon material, tip, markerbands and catheter.This failure is likely due to an interaction between the user and device, or sample, that caused or contributed to the error.It is likely that the failure may be related to unintended inappropriate use of the device and incorrect sample preparation.Therefore, the most probable root cause is unintended use error caused or contributed to event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the directions for use (dfu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a uromax ultra dilation balloon catheter was used in the ureter during a right ureteroscopy (urs) procedure performed on (b)(6) 2020.According to the complainant, during preparation, a leak was noted on the balloon.The procedure was completed with another uromax ultra dilation balloon catheter.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UROMAX ULTRA
Common Device Name
DILATOR, CATHETER, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9922396
Report Number3005099803-2020-01491
Device Sequence Number19651454
Product Code KOE
UDI-Device Identifier08714729341116
UDI-Public08714729341116
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K130804
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model NumberM0062251010
Device Catalogue Number225-101
Device Lot Number0024038801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2020
Initial Date Received by Manufacturer 03/13/2020
Supplement Date Received by Manufacturer04/24/2020
Initial Report FDA Received Date04/03/2020
Supplement Report FDA Received Date05/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
-
-