• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PUNCTUA ICD; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PUNCTUA ICD; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number E051
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The returned implantable pulse generator was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to this anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was not included in the advisory population.
 
Event or Problem Description
It was reported that this device declared a fault code-1003 indicative to voltage too low for remaining capacity.This device was explanted at a surgical intervention and a new device was implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA ICD
Common Device Name
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9922611
Report Number2124215-2020-03670
Device Sequence Number3296027
Product Code LWP
UDI-Device Identifier00802526480683
UDI-Public00802526480683
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S235
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/16/2014
Device Model NumberE051
Device Catalogue NumberE051
Device Lot Number100533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2020
Initial Report FDA Received Date04/03/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age66 YR
-
-