DEPUY IRELAND - 9616671 ACTIS BROACH SZ 7; BROACHES Back to Search Results
Model Number 2010-01-070
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/04/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.The root cause is attributed to misuse.Deep circular scratching is due to the taper not being seated all the way into the broach handle before locking down on the lever locking mechanism of the broach handle.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the size 7 actis broach was damaged during normal use and needs to be replaces.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 7
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key9924421
MDR Text Key186561880
Report Number1818910-2020-10201
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393696
UDI-Public10603295393696
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2010-01-070
Device Catalogue Number201001070
Device Lot NumberPG288690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2020
Initial Date Manufacturer Received 03/24/2020
Initial Date FDA Received04/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age69 YR
-
-