• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GENTHERM MEDICAL, LLC HEMOTHERM; HEATER COOLER DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GENTHERM MEDICAL, LLC HEMOTHERM; HEATER COOLER DEVICE Back to Search Results
Model Number 400CE
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Customer reported the unit started draining during a procedure.No injury reported.Device received and evaluated: debris was found in heat outlet solenoid and the power cord was altered.Both issues were corrected.
 
Event or Problem Description
Customer reported issue: during a procedure while in heating mode the heat reservoir began to empty.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM
Common Device Name
HEATER COOLER DEVICE
Manufacturer (Section D)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati, oh
MDR Report Key9924829
Report Number1516825-2020-00004
Device Sequence Number14155150
Product Code DWC
UDI-Device Identifier10613031860221
UDI-Public(01)10613031860221
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K122813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2020
Initial Report FDA Received Date04/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-