| Model Number |
400CE |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer reported the unit started draining during a procedure.No injury reported.Device received and evaluated: debris was found in heat outlet solenoid and the power cord was altered.Both issues were corrected.
|
| |
|
Event or Problem Description
|
|
Customer reported issue: during a procedure while in heating mode the heat reservoir began to empty.
|
| |
|