| Model Number |
0128 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated should different information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was an attempted implant due to high out of range pacing impedance measurements greater than 2,000 ohms.The lead was returned.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was an attempted implant due to high out of range pacing impedance measurements greater than 2,000 ohms.The lead was returned.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|