RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 03apr2020.
|
|
Event Description
|
It was reported that the unit had a button failure on the navigation ring.The device was in use at the time of the event; however, there was no patient harm.The manufacturer's field service engineer (fse) performed troubleshooting and confirmed the reported issue.The fse replaced the user interface assembly, navigation ring button and switch and the issue was resolved.
|
|
Manufacturer Narrative
|
G4: 08apr2020.B4: 10apr2020.H11: h6: correction to the patient code: there was no patient involvement.The unit was not in clinical use when the reported event occurred.The issue was discovered during preventative maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4: 24sep2020.B4: 29sep2020.D4: udi#: (b)(4).A user interface (ui) assembly was returned for analysis.A visual inspection of the returned component was performed, and no notable conditions were found.The returned component was installed into a test ventilator for analysis.An investigation was performed and the product analysis technician reported that no failures were identified with the returned component.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|