| Model Number |
MMT-551LNAB |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
03/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was 500 mg/dl.The customer stated that the crack on back of the insulin pump, screen and battery compartment.The customer stated that the reservoir was able to lock into place.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|