• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22201A
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transcervical resection (tcr) procedure, sparks occurred during coagulation and it was then noticed that the loop wire at the distal end of the hf resection electrode had broke off and fallen into the patient.An x-ray examination was performed but the broken off fragment could not be located.No further information was provided but the intended procedure was completed with the same set of equipment and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, based on the customer's description and our experience, the reported damage can most likely be attributed to use-related wear and tear.Since the user facility could not definitely determine the lot number of the relevant hf resection electrode, a manufacturing and quality control review was performed for all possible lot numbers (lot #: 13303p03l001, device manufacturer date: 05.11.2013; lot #: 14057p03l003, device manufacturer date: 11.03.2014; lot #: 14058p03l001, device manufacturer date: 07.03.2014) of the resection electrode without showing any abnormalities related to function and safety.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12
Common Device Name
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key9927693
Report Number9610773-2020-00100
Device Sequence Number2764184
Product Code GCP
UDI-Device Identifier04042761021302
UDI-Public04042761021302
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22201A
Device Catalogue NumberA22201A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/19/2020
Supplement Date Received by Manufacturer04/21/2020
Initial Report FDA Received Date04/06/2020
Supplement Report FDA Received Date05/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS UES-40 SURGMASTER ELECTROSURGICAL UNIT
-
-