Model Number 6L27-25
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction/removal number= 3002809144-03/16/20-002-r investigation into the issue confirmed a performance shift for the architect havab-g impacted reagent lots due to the erroneous concentration for (b)(6) virus that was utilized during manufacture, which has the potential to generate (b)(6) elevated control and patient sample results.An internal study using (b)(6) patient samples was conducted and determined that results that fall in the range of (b)(6) have the potential to be (b)(6).A product recall letter has been issued to all customers who have received the impacted lots 03429be00, 06172be00, 08073be00, and 10353be00.The product recall letter instructs the customer to immediately discontinue the use of, and destroy, any remaining inventory of the specific 4- architect havab-g reagent lots and instructs the customer to contact customer support for replacement material in the event if they were currently using or have inventory of one of the impacted lots.
 
Event or Problem Description
The customer reported that the negative quality controls were shifting high when using the architect havab-g reagent.The customer stated that the (b)(6) controls was generating results of (b)(6).There was no patient involvement and no reported impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT HAVAB-G
Common Device Name
IGG ANTI-HAV
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key9928192
Report Number3002809144-2020-00301
Device Sequence Number14157616
Product Code LOL
UDI-Device Identifier00380740011185
UDI-Public00380740011185
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K113704
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2020
Device Model Number6L27-25
Device Catalogue Number06L27-25
Device Lot Number03429BE00
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2020
Initial Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3002809144-03/16/20-002-
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I1000SR ANALYZER, LN 01L86-01,; SN (B)(4).
-
-