• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ1 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ1 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-12-070
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191)
Date of Event 04/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had increasing hip & groin pain.Head & liner exchange.After review of medical records, it was indicated that patient presents with progressive weight-bearing pain and some intermittent mechanical symptoms of catching or ratcheting in her right hip.Her evaluation revealed that she did have elevated cobalt and chromium ion levels, and an mri did show a moderate-sized periarticular fluid collection.The patient was then revised for failed right total hip arthroplasty due to metallosis and local adverse tissue reaction as well as early periacetabular osteolysis.Operative notes reported that there was blackish and grayish metallosis staining within the synovial lining diffusely of the hip joint, and there was some wear debris within the capsular structures.There was clear trunnionosis with wear debris at the junction of the femoral head and the trunnion.There was some mild yellowish wear debris on the posterior aspect of the liner, surgeon felt the majority of the metallosis wear debris was from the trunnion.There was evidence for some osteolysis.Doi: (b)(6) 2007; dor: (b)(6) 2019; right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ1 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9928393
Report Number1818910-2020-10231
Device Sequence Number2818451
Product Code LPH
UDI-Device Identifier10603295060123
UDI-Public10603295060123
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K011489
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1570-12-070
Device Catalogue Number157012070
Device Lot NumberBY1A81000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/11/2020
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date04/06/2020
Supplement Report FDA Received Date10/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight63
-
-