Model Number 1550300-33
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned to abbott vascular for analysis.Return of the stent implant may have further aided the analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot and relabeled lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a mid left anterior descending artery.After confirming the lesion with an optical coherence tomography, a 3.0x33mm xience sierra stent was advanced to the lesion but was noted as too short to cover the lesion.Therefore, a 38mm length xience stent was used instead to successfully complete the procedure.There were no adverse patient effects or clinically significant delay.No additional information was provided.3/16/2020: additional information received from the account where the physician states uncertainty whether the stent was undersized than expected, or became shorter during the procedure.There were no adverse patient effects or clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9929122
Report Number2024168-2020-03403
Device Sequence Number2717799
Product Code NIQ
UDI-Device Identifier08717648227295
UDI-Public08717648227295
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/04/2021
Device Model Number1550300-33
Device Catalogue Number1550300-33
Device Lot Number8032741
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2020
Initial Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-