| Model Number |
1506-00-002 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); No Code Available (3191)
|
| Date of Event |
06/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient received an attune total knee to treat left knee osteoarthritis.The patella was resurfaced and depuy cement x 2 was utilized.The procedure was completed without complications.Patient received a left knee revision to treat pain secondary to tibial tray loosening.Intraoperative findings confirmed loosening of the tibial tray at the cement to implant interface.The femoral component and patella were well-fixed and retained.There was no reported product problem with the explanted tibial insert.The patient was revised with attune products utilizing competitor cement.The procedure was completed without complications.Doi: (b)(6) 2017.Dor: (b)(6) 2019; left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|