Model Number 1506-40-007
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 01/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Initial reporter occupation: (b)(6).(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient received a total left tka revision to treat persistent pain.The tibial tray was loose and completely debonded at the cement to implant interface.The surgeon notes that the soft tissue surrounding the tibial cement mantle was necrotic and needed extensive debridement.The procedure was completed without complications.Doi: (b)(6) 2015.Dor: (b)(6) 2018 (tray and insert (b)(4)).Dor: (b)(6) 2020 (total revision).Left knee.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUNE CRS FB TIB BASE SZ 7 CEM
Common Device Name
ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9929311
Report Number1818910-2020-10267
Device Sequence Number2416458
Product Code JWH
UDI-Device Identifier10603295042815
UDI-Public10603295042815
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-40-007
Device Catalogue Number150640007
Device Lot Number8634108
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/23/2020
Supplement Date Received by Manufacturer05/08/2020
Initial Report FDA Received Date04/06/2020
Supplement Report FDA Received Date05/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CEMENTED STEM 14X50MM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 8 16MM; ATTUNE PS FEM LT SZ 8 CEM; COMPETITOR BONE CEMENT; ATTUNE CEMENTED STEM 14X50MM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 8 16MM; ATTUNE PS FEM LT SZ 8 CEM; COMPETITOR BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
-
-