Model Number 04.037.161S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Arthritis (1723)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent removal of a proximal femoral nailing system (tfna) to convert to a total hip arthroplasty due to post-traumatic arthritis.The patient's status is unknown.This complaint involves three (3) devices.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: investigation summary: the complaint condition of an post-traumatic arthritis could be confirmed, but there is no product problem reported.Product was not returned.Based on the information available, it has been determined that no corrective and preventive action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H3, h4, h6: a review of the device history record.Device history lot manufacturing location: monument, manufacturing date: 14-dec-2017, expiration date: 30-nov-2027, part number: 04.037.161s, 11mm/130 deg ti cann tfna 400mm / left ¿ sterile, lot number: h523658 (sterile) , lot quantity: 5.One piece was scrapped in cell at op #50, mill flutes and relief, due to a raw material cosmetic defect.Work order traveler met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, in process / inspect dimensional / final, met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, tfna assembly inspection, met all inspection acceptance criteria.Packaging label log lppf, lmd/lpf rev d was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 14464 supplied by ees (albuquerque) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removed due to post-traumatic arthritis¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Device history batch null, device history review 19-mar-2020: dhr reviewed.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11MM/130 DEG TI CANN TFNA 400MM/LEFT - STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9929423
Report Number2939274-2020-01730
Device Sequence Number2818354
Product Code HSB
UDI-Device Identifier10886982096692
UDI-Public(01)10886982096692
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.037.161S
Device Catalogue Number04.037.161S
Device Lot NumberH523658
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/10/2020
Supplement Date Received by Manufacturer04/09/2020
Initial Report FDA Received Date04/06/2020
Supplement Report FDA Received Date04/28/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
11MM/130 DEG TI CANN TFNA 400MM/LEFT - STERILE; TFNA FENESTRATED HELICAL BLADE 100MM; UNK - SCREWS: LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
Patient Weight106
-
-