• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS; GAUGE,DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS; GAUGE,DEPTH Back to Search Results
Model Number 357.402
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Part number: 357.402, synthes lot number: 4695608, supplier lot number: n/a, release to warehouse date: 17dec2003, expiration date: n/a.Manufactured by synthes brandywine.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Visual inspection: the instrument was inspected, and the ball ø2.5 mm assembly and compression spring are missing from the construct preventing the device to function as intended.Document/specification review the following drawing(s) was reviewed: - locking bolt measuring device.- slider assembly.No design or manufacturing defect or deficiency was observed during the investigation.From the drawings in can be seen that the ball ø2.5 mm and compression spring are components of the instrument and therefore a conclusive determination has been reached.The overall complaint was not confirmed for the instrument as the device is missing the ball and spring components and is not broken.A definitive root cause for this complaint could not be determined.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the two (2) locking bolt measuring device for trochanteric fixation nails were broken during reverse logistics audit of returned device at millstone.There was no patient involvement and no additional information is available.This report involves one (1) locking bolt measuring device f/trochanteric fixation nails.This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING BOLT MEASURING DEVICE F/TROCHANTERIC FIXATION NAILS
Common Device Name
GAUGE,DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9930313
Report Number2939274-2020-01735
Device Sequence Number19436900
Product Code HTJ
UDI-Device Identifier10886982196255
UDI-Public(01)10886982196255
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number357.402
Device Catalogue Number357.402
Device Lot Number4695608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2020
Initial Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-