PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problem Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the mobile programmer application froze during an implant procedure.The user indicates the programmer stayed pacing at the programmed rate of 80 beat per minute (bpm) and an output of 5 volts at 0.5 milliseconds.The user was unable to program any other parameters as all other options appeared grayed out on the screen and an alternative programmer was used to complete the procedure.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: analysis of the device logs was able to confirm the customer comment of a mobile programmer application freeze.A total of 4 errors were noted from the device logs.There are 2 short telemetry loss situations noted, which is theoretically not possible due the pacing system analyzer telemetry connection implementation.The issue was observed at the phase, when multiple downlink requests carrying the programmed parameters failed due to the no response from the pacing system analyzer device (the request was delivered there but the pacing system analyzer didn't respond in 10 seconds).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC BASE
Type of Device
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key9930534
MDR Text Key187084739
Report Number3004593495-2020-00395
Device Sequence Number1
Product Code DTC
UDI-Device Identifier00763000201289
UDI-Public00763000201289
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Date Manufacturer Received04/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age81 YR
-
-