• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC ARTIC-L 3D TI SPINAL SYSTEM WITH TIONIC TECHNOLOGY; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC ARTIC-L 3D TI SPINAL SYSTEM WITH TIONIC TECHNOLOGY; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 56250905
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented with lumbar spinal canal stenosis; and underwent transforaminal lumbar interbody fusion at l3-s1.On an unknown date, post-op, the cage (that was implanted at l3-l4) backed out from the intervertebral disc space.Additionally, on both the left side and the right side, the rods had deviated from the screws at l3 to the caudal side, but the set screws remained fitted on the screws.Reportedly, although, a revision surgery had been scheduled for (b)(6) 2020 due to posterior junctional kyphosis, the patient did not want to undergo a revision surgery.So, the revision surgery would not be performed currently.The surgeon suspected that the compression loosened due to the rod deviated, which led to the result that the cage backed out.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A revision surgery was performed, in which extension of fusion was performed.As "ai" backed out totally and came off to the posterior side, "dd" was placed at l3-l4 from the anterior side to the intervertebral disc space.Additional screw insertion was performed at l2.The screw at l3, from which the rod came off, was also replaced.Connection was performed using connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIC-L 3D TI SPINAL SYSTEM WITH TIONIC TECHNOLOGY
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key9933939
Report Number1030489-2020-00429
Device Sequence Number16332326
Product Code MAX
UDI-Device Identifier00763000085438
UDI-Public00763000085438
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171689
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number56250905
Device Lot NumberCA18J171
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/18/2020
Supplement Date Received by Manufacturer05/03/2020
Initial Report FDA Received Date04/08/2020
Supplement Report FDA Received Date06/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RODS; RODS
Outcome Attributed to Adverse Event Required Intervention;
-
-