ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550300-08
Medical Device Problem Codes Difficult to Insert (1316); Material Separation (1562); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.The reported material separation was able to be confirmed.The reported difficult to insert and the reported deformation due to compressive stress were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure.Further manipulation resulted in the noted stretched tip and ultimately resulted in the reported tip separation; thus resulting in the reported difficulty to insert.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat an unspecified lesion.A 3.0x08mm xience sierra had difficulty being inserted onto the guide wire.The tip was noted as bent then broke off.Another same size xience sierra was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delays.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9934340
Report Number2024168-2020-03455
Device Sequence Number16318929
Product Code NIQ
UDI-Device Identifier08717648227233
UDI-Public08717648227233
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/03/2020
Device Model Number1550300-08
Device Catalogue Number1550300-08
Device Lot Number9112741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2020
Initial Report FDA Received Date04/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-