| Model Number |
MMT-522LNAB |
| Medical Device Problem Codes |
Visual Prompts will not Clear (2281); Application Program Version or Upgrade Problem (2881); Loss of Data (2903)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/31/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had multiple pump error alarm and frozen display.No harm requiring medical intervention was reported.The customer was assisted with troubleshooting.Customer received a second alarm after clearing the alarm.Customer reported they were unable to clear the alarm.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received alarming a32 followed by an e22 alarm, problem isolated to mother board.Unable to perform operating currents, self test and displacement test or verify frozen screen due to a32 and e22 alarm.
|
| |
|