| Model Number |
MMT-722NAB |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had compromised broken force sensor alarm during priming.Customer reported that the drive support cap was loosened.During seating the force sensor did not detect the reservoir.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test.However device alarmed motor error during basic occlusion test due to loose or protruded drive support disk.Unable to perform occlusion test, prime/a33 test, excessive no delivery test or verify a33 alarm due to motor error.
|
| |
|