MAUDE Adverse Event Report: INTERVASCULAR SAS HEMAGARD CAROTID PATCH KNITTED ULTRATHIN; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS HEMAGARD CAROTID PATCH KNITTED ULTRATHIN; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HGKTP08/75CPUT(1)
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The review of historical data indicated that no other similar complaint was reported for the same sterilization lot number.The device history records review concludes that there is no non-conformance / planned deviation in relation with the event reported.It was reported that the product is available for investigation, it should be returned to intervascular for examination.The investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
(b)(4).The customer ordered a patch hgktp08/75cput (1).The patch is the utrathin, the box is from a knitted and the sticker is from an ultrathin.
 
Event or Problem Description
See initial mfr report#1640201-2020-00009.Complaint # (b)(4).
 
Additional Manufacturer Narrative
(10/3221) no investigation could be performed on the actual device.Indeed, the device involved in the event is available but not received by manufacturer at this date.The delay in returning the product is due to the current sanitary crisis.However, since the problem was a mismatch between box and product, we were able to investigate the issue with the use of pictures provided by the hospital.See codes 4112/4245 below.(4112/4245) the qa manager carried out an investigation based on pictures provided by the hospital and our traceability data.The main conclusions are: the most probable origin of the event is an error occurring during a defective box replacement taking place either during the preparation of the packaging components or either during the final packaging process itself.No other product has been impacted.(25) the conducted investigation concluded that the origin of the event is a punctual error occurring during the manufacturing process and confirmed that the product does not conform to the specification.Therefore, a non-conformity report has been initiated in order to investigate the root cause and take appropriate corrective actions if necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD CAROTID PATCH KNITTED ULTRATHIN
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat
MDR Report Key9937990
Report Number1640201-2020-00009
Device Sequence Number2496995
Product Code DSY
UDI-Device Identifier00384401015208
UDI-Public00384401015208
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K983819
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberHGKTP08/75CPUT(1)
Device Catalogue NumberHGKTP08/75CPUT (1)
Device Lot Number20B20
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/10/2020
Supplement Date Received by Manufacturer05/05/2020
Initial Report FDA Received Date04/08/2020
Supplement Report FDA Received Date06/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-